Setup option
The setup option is used to select the thermocouple type that matches the type that will be
connected to the meter, to switch the sleep mode OFF/ON, and to adjust the thermocouple offset.
Note: In the setup mode, the upper display shows the menu items and the lower display shows
the selection for a given menu item.
Press and hold the SET button for 3 seconds to enter or exit the Setup mode. When the
thermometer is in Setup mode, the display shows the SETUP display icon.
Set Type J or K thermocouple
1. Press ▲ or ▼to scroll to the “TYPE” setting (shown on the bottom left of the display).
2. Press ENTER to choose the option.
3. Press ▼ to toggle between J and K.
4. Press and hold the SET button for 3 seconds to save the setting and return to normal mode.
Set Sleep/Non-sleep mode
The thermometer enters sleep mode by default. The meter will automatically shut off after 20
minutes of inactivity (no buttons pressed). When the thermometer is in Setup mode, the display
shows the SETUP display icon.
1. Press ▲ or ▼to scroll to the SLP setting.
2. Press ENTER to choose the option;
On or OFF will appear in the display.
3. Use the arrow key ▼ to select ON or OFF.
4. Press ENTER to store the new setting.
5. Press and hold the SET button for 3 seconds to return to normal mode.
Set Offset
Use the offset option in Setup (T1 menu item) to adjust the thermometer’s readings to
compensate for known temperature offset at a specific temperature. The adjustment range is ±
5.0°C or ± 9.0 °F.
1. Plug the thermocouple into the input connector.
2. Place the thermocouple into the known, stable temperature environment (such as an ice
bath or dry well calibrator). Allow the reading to stabilize.
3. Use the arrow keys ▲ ▼ to scroll to the “T1” option and then press ENTER.
4. Change the offset temperature reading (upper display) by pressing ▼until the desired
temperature is reached.
5. Press and hold the SET button for 3 seconds to return to normal mode.