5
Specication
Electrical Requirements 115VAC/60Hz
Power Consumption 75W
Dimensions 4.1” (105mm) High
×5.7” (145mm) Wide
× 7.9” (200mm) Deep
Weight 1.6kg (3.53lbs)
Sound Pressure Level 58dB(A)
Max. Comp. Pressure 30 Psig or greater
Neb. Operating Pressure 10~14 Psig
Liter Flow Range 4 - 8 LPM
Particle Size Less Than 5 microns
Operating Temperature 10
o
C ~ 40
o
C (50
o
F to 104
o
F)
Operating Humidity Up to 95% non-condensing
Storage/Transport Temperature Range -20
o
C ~ 60
o
C (-4
o
F to 140
o
F)
Storage/Transport Humidity Up to 95% non-condensing
(1). Conditions may vary based on altitude above sea level, barometric pressure and
temperature. Specication subjects to change without notice.
(2). This product complies with the following standards: C22.2 No 601.1-M90, UL Std.
No 2601-1 and IEC/EN 60601-1-1.
(3). For electromagnetic compatibility the standards are based as follow: CSA 601.1-2
and IEC/EN 60601-1-2.
(4). Protection against electric shock is classied as Type II.
(5). Degree of protection against electric shock is classied as Type BF.
(6). Mode of operation belongs to continuous mode.
(7). Equipment not suitable for use in the presence of a ammable anesthetic mixture
with air or with oxygen or nitrous oxide
(8). Degree of protection against ingress of water is rated as IPX0.
Warning: To reduce the risk of burns, electrocution, re or injury:
Warning: To reduce the risk of infection: