11
What to do if ...
Problem Reason Solution
(flashing)
Batteries are flat.
Memory readings can be
called up.
Measurements are not
possible.
Insert new batteries.
e.g. Error codes 01-29:
Device may not be correctly
positioned on the wrist.
Refit device and repeat measurement
as shown and described under
«Measuring».
e.g. Error codes 30 and
higher:
System error
Repeat measurement. If problem
persists, have device checked at a
Braun Service Centre.
Measurement or
inflation does not
start instantly.
Right measuring position
was not reached.
Repeat measurement as shown and
described under «Measuring» or de-
activate the positioning system (A).
Readings from
repeated meas-
urements differ
considerably.
Blood pressure is a
fluctuating value.
For healthy adults,
deviations of about 10 to
20 mmHg are possible.
none
Cuff is not properly fitted.
Ensure that the appliance is fitted
as follows:
– on top of the slim side of the forearm
– with the display on top
– with the battery compartment facing
the user’s body
– approx. 2 cm away from the thumb root
Readings were not con-
sistently taken at heart
level.
For each measurement, proceed
exactly as shown and described under
«Measuring».
Talking, coughing,
laughing, moving etc.
when measuring will
influence the reading.
While measuring, relax, keep still,
do not move, or speak.
Prior to a mea-
surement, the
positioning
system (A) on the
display does not
work.
Positioning system (A) is
deactivated.
Refer to the chapter «Select mode»
to activate the positioning system
(A).
................
................
6083402_BP3510_S4-54 Seite 11 Dienstag, 16. Januar 2007 1:18 13