Braun 6013 Thermometer User Manual


 
carrying Braun ThermoScan thermometers or at Braun Service
Centers.
If you have accidentally placed the thermometer in your ear with-
out a lens filter in place, be sure and clean the probe (see
«Care and cleaning» and then apply a clean lens filter.
Care and cleaning
The probe tip is the most delicate part of the thermometer.
It has to be clean and intact to ensure accurate readings.
If the thermometer is ever accidentally used without a lens filter,
clean the lens as follows:
Hold the appliance with the probe tip facing down to prevent
liquid from entering the probe tip area. Very gently wipe the sur-
face with a cotton swab or soft cloth moistened with alcohol.
After cleaning, allow at least 45 minutes drying time before reat-
taching a new, clean lens filter and taking a temperature.
If the probe tip is damaged, contact Braun.
Use a soft, dry cloth to clean the thermometer display and exterior.
Do not use abrasive cleaners. Never submerge the thermometer in
water or any other liquid.
Store thermometer and lens filters in a dry location free from dust
and contamination and away from direct sunlight. The ambient
temperature at the storage location should remain fairly constant
and within the range of 50 °F to 104 °F (10 °C - 40 °C).
Always keep cleaning solutions and rubbing alcohol away
from children.
Replacing the batteries
The thermometer is supplied with two 3 V lithium batteries
(CR 2032/DL 2032). Insert new batteries when the battery symbol
appears on the display.
Using the tip of a ball-point pen, press the battery door lock to
open the battery compartment. Remove the batteries and replace
with new batteries, making sure the poles are in the right direction.
Slide battery door back in until it snaps in place.
11
6012136_IRT_3000_NA 28.10.2005 8:20 Uhr Seite 11