A&D UC-324TMW Scale User Manual


 
A&D Medical (Toll-Free): 1-888-726-9966 | Page E-7
Before Requesting Service
Problem Recommended Action
Nothing appears on
the display, even when
knocking the scale or
pressing the switches.
• Reinstall the batteries with negative
and positive terminals matching those
indicated on the battery compartment.
• Replace all batteries with new ones.
E-0 or E- s displayed.
• Is anything touching the scale?
• Is there anything under the scale?
• Remove and reinstall the batteries. Use
new ones.
888.8 is displayed and the
scale is turned off.
There is a large difference
between actual weight
and measured weight
• Place the scale on a solid, firm, and
level surface.
• Did you allow the scale to zero before
applying weight?
• Is there anything under the scale?
ActiLink/computer software
is not receiving new
measurements
• Is your computer on and your ActiLink
plugged in?
• Move the scale closer to the ActiLink
with a clear line-of-sight and take a
measurement.
• Replace all batteries with new ones.
H is displayed
The 330lb (150kg) weight capacity has
been exceeded. Please reduce weight
and re-weigh.
L is displayed
Re-weigh and wait until weight is
displayed on the screen and/or a final
beep is heard.
Note: If the actions described above do not solve the problem, please
contact A&D Medical. Do not attempt to open or repair this product, as any
attempt to do so will make your warranty invalid.
Maintaining the Scale
Clean the scale with a soft dry cloth.
Never use water, detergent, alcohol, benzene, thinner or other harsh
chemicals to clean the scale.
Remove the batteries if the scale is not going to be used for a long
time.