A&D SW-150KM Scale User Manual


 
33
Five-level comparator modeSet the HIHI limit value.
Three-level comparator modeSet the HI limit value.
(No reply for the function setting “aCk 0 “.)
Set the six-digit value excluding the polarity and decimal point.
Command
H 2
,
+ 0 0 0400C
R
L
F
Reply
H 2
,
+ 0 0 0 4 0 0C
R
L
F
Five-level comparator modeSet the HI limit value.
Three-level comparator modeNot used.
(No reply for the function setting “aCk 0 “.)
Send the six-digit value excluding the polarity and decimal point.
Command
H 1
,
+ 0 0 0300C
R
L
F
Reply
H 1
,
+ 0 0 0300C
R
L
F
Five-level comparator modeSet the LO limit value.
Three-level comparator modeNot used.
(No reply for the function setting “aCk 0 “.)
Send the six-digit value excluding the polarity and decimal point.
Command
L 1
,
+ 0 0 0200C
R
L
F
Reply
L 1
,
+ 0 0 0200C
R
L
F
Five-level comparator modeSet the LOLO limit value.
Three-level comparator modeSet the LO limit value.
(No reply for the function setting “aCk 0 “.)
Send the six-digit value excluding the polarity and decimal point.
Command
L 2
,
+ 0 0 0100C
R
L
F
Reply
L 2
,
+ 0 0 0100C
R
L
F
Replies to the commands other than examples above when the function setting “aCk 1 “ is
selected.
The scale is not in a state where a command could be executed. Then, the scale will reply “I”.
Command
Z C
R
L
F
Reply
I C
R
L
F
The scale is not in a condition that zero operation is possible.
Command does not exist for the scale. Then, the scale will reply “?”.
Command
B C
R
L
F
Reply
? C
R
L
F
The scale received an undefined command.
When the function setting “aCk 0 “ is selected, undefined commands are ignored
and no reply is sent.