A&D EJ-120 Scale User Manual


 
36
12. SPECIFICATIONS
12-1. EJ series
MODEL EJ-120 EJ-200 EJ-300 EJ-410 EJ-610
Weight capacity 120 g 210 g 310 g 410 g 610 g
Min. display “d” 0.01 g 0.01 g 0.01 g 0.01 g 0.01 g
Repeatability
(Std. deviation)
0.01 g 0.01 g 0.01 g 0.01 g 0.01 g
Linearity ±0.01 g ±0.01 g ±0.02 g ±0.02 g ±0.02 g
Sensitivity drift ±20 ppm / °C (10°C~30°C / 50°F~86°F)
No. of samples 5, 10, 25, 50 or 100 pieces
Max. count * 12,000 pcs 21,000 pcs 31,000 pcs 41,000 pcs 61,000 pcs
Min. unit weight * 0.01 g
Min. % display 0.1 %
Min. 100 % weight 1 g
Display 7 segment LCD display with backlight (Character height 16 mm)
Display update 10 time per second
Operating temp. -10°C~40°C / 14°F~104°F, less than 85% R.H. (non-condensing)
Power supply AC adapter or 4 x “AA” size dry-cell batteries
Battery operation Approximately 70 hours (backlight off, alkaline batteries)
Weighing pan size 110 mm ø
Weight
Approximately 850 g
Calibration weight
(factory setting)
100 g 200 g 300 g 400 g 600 g
* In case of “Umin 0” (factory setting)
MODEL EJ-1500 EJ-2000 EJ-3000 EJ-4100 EJ-6100
Weight capacity 1500 g 2100 g 3100 g 4100 g 6100 g
Min. display “d” 0.1 g 0.1 g 0.1 g 0.1 g 0.1 g
Repeatability
(Std. deviation)
0.1 g 0.1 g 0.1 g 0.1 g 0.1 g
Linearity ±0.1 g ±0.1 g ±0.2 g ±0.2 g ±0.2 g
Sensitivity drift ±20 ppm / °C (10°C~30°C / 50°F~86°F)
No. of samples 5, 10, 25, 50 or 100 pieces
Max. count * 15,000 pcs 21,000 pcs 31,000 pcs 41,000 pcs 61,000 pcs
Min. unit weight * 0.1 g
Min. % display 0.1 %
Min. 100 % weight 10 g
Display 7 segment LCD display with backlight (Character height 16 mm)
Display update 10 time per second
Operating temp. -10°C~40°C / 14°F~104°F, less than 85% R.H. (non-condensing)
Power supply AC adapter or 4 x “AA” size dry-cell batteries
Battery operation Approximately 70 hours (backlight off, alkaline batteries)
Weighing pan size 127 mm x 140 mm
Weight
Approximately 970 g
Calibration weight
(factory setting)
1500 g 2000 g 3000 g 4000 g 6000 g
* In case of “Umin 0” (factory setting)