34
way, or if the controller is switched off before the last step, no
calibration data is stored in non-volatile memory (EEPROM).
ANALOG OUTPUT CALIBRATION
In the meters where the analog output is available, this fea-
ture is factory calibrated through software. The user may also
perform these calibration procedures.
IMPORTANT It is recommended to perform the output calibration at least
once a year. Calibration should only be performed after 10
minutes from power up.
• With a multimeter or an HI 931002
connect the common port to the
ground output and the second port
to the current output or voltage out-
put (depending on which parameter
is being calibrated).
• Press and hold in sequence CFM first, then and then
CAL to enter the Analog Output Calibration mode.
• Execute the password procedure.
• The primary display will show the current selected param-
eter blinking. Use the to select the code (0-5 see chart
below) for the desired parameter displayed on the second-
ary display (e.g. 4-20 mA).
• Press CFM to confirm the selected parameter that will stop
blinking on the primary display. The secondary display shows
the HI 931002 or multimeter input value as lower limit of
the interval.