Code Valid Values Default PW
14 Relay 1 deviation (D1) 0.5 to 10% f.s. 1% f.s. no
15 Relay 1 reset time 0.1 to 999.9 minutes 999.9 no
16 Relay 1 rate time 0.0 to 999.9 minutes 0.0 no
21 Relay 2 mode (M2) same as relay 1 0 no
22 Relay 2 setpoint (S2) 0.5 to 99.5% full scale 75% f.s. no
23 Relay 2 hysteresis (H2) 0 to 5% f.s. 1% f.s. no
24 Relay 2 deviation (D2) 0.5 to 10% f.s. 1% f.s. no
25 Relay 2 reset time 0.1 to 999.9 minutes 999.9 no
26 Relay 2 rate time 0.0 to 999.9 minutes 0.0 no
30 Relay 3 high alarm (HA) 0.5 to 99.5% full scale 95% f.s. no
HA-Hys>LA+Hys,Hys=1.5%f.s.,HA>S1 or HA>S2
31 Relay 3 low alarm (LA) 0.5 to 99.5% full scale 5% f.s. no
LA+Hys<HA-Hys,Hys=1.5%f.s.,LA<S1 or LA<S2
32 Proportional control 1 to 30 min 5 no
mode period
33 Maximum relay ON time 1 to 10 min 10 no
(after which an alarm mode is entered)
34 Alarm mask time 00:00 to 30:00 00:00 no
40 Analog output selection 0: 0-1mA 2 no
1: 0-20 mA
2: 4-20 mA
3: 0-5 VDC
4: 1-5 VDC
5: 0-10 VDC
41 Analog output 0 to 100% full scale 0 no
lower limit (O_VARMIN) (O_VARMIN < O_VARMAX - 5% f.s.)
42 Analog output 0 to 100% full scale 100% f.s. no
upper limit (O_VARMAX) (O_VARMIN < O_VARMAX - 5% f.s.)