I have lost my manual and I need to know what do I do when E-5 comes up on my screen - I have misplaced my manual. Thanks P. Sims

Asked by Peggy on 03/11/2009 1  Answer

ManualsOnline posted an answer 15 years, 1 month ago

The ManualsOnline team has found the manual for this product! We hope it helps solve your problem. Get the Manual Here
4 Meaning O The measuring window is dirty. Press the Test button to turn the meter off. Clean the measuring window (see Chapter 8.1) and repeat the test with a new test strip. O A test strip is stuck in the meter. Hold the meter upright so the test strip is pointing downwards. Press the Test button to turn the meter off and allow the test strip to drop out. If the test strip does not drop out, remove it manually, even if you feel some resistance. O The test strip drew up too little blood or control solution. Turn the meter off and then on again by pressing the Test button. Repeat the test with a new test strip and a larger amount of blood or control solution.
Was this answer helpful?

Add Your Answer