Operation Manual
Important: Do not operate the
John Bunn DigiOx Finger Pulse
Oximeter without rst reading
and understanding this manual!
Save this manual for future use.
Note: The most current version of this manual
can be found online at www.grahameld.com
DigiOx
Finger Pulse Oximeter
JB02008-INS-LAB-RevA09
CONTENTS
INTRODUCTION ............................................................3
INTENDED USE
.......................................................3
CONTRAINDICATIONS
............................................3
IMPORTANT SAFETY GUIDELINES - PLEASE
READ BEFORE USE
............................................3
OVERVIEW
....................................................................7
APPEARANCE
..............................................................7
FEATURES
....................................................................7
NAME AND MODEL
......................................................8
BATTERY INSTALLATION
............................................8
OPERATION INSTRUCTIONS
......................................8
ACCESSORIES (INCLUDED)
.....................................11
REPAIR, MAINTENANCE, AND DISPOSAL
...............11
MAINTENANCE
......................................................11
CLEANING AND DISINFECTION
..........................12
DISPOSAL
..............................................................13
TROUBLESHOOTING
.................................................13
SYMBOL KEY
..............................................................15
TECHNICAL SPECIFICATIONS
..................................16
LIMITED WARRANTY
.................................................18
MANUAL DE OPERACIÓN (ESPAÑOL)
.....................21
INTRODUCTION
Dear User: Please read and understand all instructions
before using the John Bunn DigiOx Finger Pulse
Oximeter, JB02008.
Intended use
The John Bunn DigiOx Finger Pulse Oximeter is used
to measure oxygen saturation (SpO
2
) and pulse rate in
home or clinical settings. It is designed for adult and
pediatric patients.
Contraindications
WARNING: The oximeter is not intended for
continuous patient monitoring. DO NOT use the
oximeter while patient is undergoing MRI or CT
scanning.
Important Safety Guidelines - Please Read Before Use
The safety statements presented in this chapter refer
to the basic safety information that the operator of the
John Bunn DigiOx Finger Pulse Oximeter shall pay
attention to and abide by. There are additional safety
statements in other chapters or sections, which may
be the same as or similar to the following, or specific
to the operations.
WARNING: Indicates a potential hazard situation
or unsafe practice that, if not avoided, could result in
death or serious injury.
Caution:
Indicates a potential hazard or unsafe
practice that, if not avoided, could result in minor
personal injury or product/property damage.
Note: Provides application recommendations or other
useful information to ensure that you get the most from
your product.
WARNINGS:
Important! Read and understand this manual before
using the oximeter. Do not use this device without
proper instruction from a Healthcare Professional.
If the oximeter is not properly operated, personal
injury and/or damage to the oximeter could result.
If components are damaged or missing, contact your
dealer immediately. DO NOT use substitute parts.
Inspect the device before use to ensure that there is
no visible damage that may affect patient or user’s
safety or measurement performance. Thereafter,
inspect the device at least once a week.
GF Health Products, Inc. assumes no responsibility
for any damage or injury caused by improper instal-
lation or use of this product.
DO NOT open the oximeter housing. It contains
no serviceable parts. Service must be performed
by trained, authorized personnel only. Otherwise,
device failure and health hazard may occur.
EXPLOSION HAZARD: Do not use the oximeter in
the presence of ammable anesthetics, explosive
substances, vapors or liquids.
The oximeter is designed for real-time and
rapid measurement. It is not suitable for long-time
continuous patient monitoring. DO NOT apply the
sensor to the same nger for more than two hours.
If any readings are found to be abnormal, please
change the position of the sensor.
The oximeter is intended only as an adjunct in
patient assessment. It must be used in conjunction
with clinical signs and symptoms. It is intended as a
measuring, not a treatment, device.
Magnetic and electrical elds are capable of interfer-
ing with the proper performance of the oximeter. For
this reason, ensure that all external devices operated
in the vicinity of the oximeter comply with the relevant
EMC requirements. Mobile phones, X-ray equipment,
and/or MRI devices are a possible source of interfer-
ence, as they may emit high levels of electromagnetic
radiation. This device complies with IEC60601-1-2.
DO NOT clip this device on edema or tender tissue.
The infrared light emitted from the device (it is invis-
ible) is harmful to the eyes. DO NOT stare into the light.
Keep this device out of the reach of children.
Cautions
To ensure patient safety, use only parts and ac-
cessories specified in this manual.
Keep the oximeter away from dust, vibration,
corrosive substances, explosive materials, high
temperature and moisture.
If the oximeter gets wet, stop operation im-
mediately and do not resume operation until it is
dry. When it is carried from a cold environment
to a warm and/or humid environment, allow it to
warm to room temperature before using it.
DO NOT use sharp objects to press the front
panel button.
DO NOT sterilize this device. Do not subject this
device to irradiation, steam, ethylene oxide, high
temperature, or immersion in liquid. Refer to the
Repair and Maintenance/Cleaning and Disinfec-
tion section of this manual for instructions for
cleaning and disinfection.
OVERVIEW
Appearance
power/display
button
battery coverdisplay screen
rubber
cushion
Figure 1 Figure 2
The DigiOx Finger Pulse Oximeter is shown above.
Features
The John Bunn DigiOx Finger Pulse Oximeter
provides a simple way to spot-check patients by
combining the sensor and monitor into one integrated,
compact, easy-to-use device with single-button
operation. The oximeter can accurately measure
oxygen saturation and pulse rate. When a finger is
inserted into the sensor’s rubber cushion, and the
Power/Display Button is pressed to turn the oximeter
on, then the SpO
2
value and pulse rate value display
on the screen. The oximeter features a bright-color
LED screen with adjustable brightness and multi-
directional displays that enable the user to read results
from any angle. The oximeter features a low voltage
indicator, powers off automatically in eight seconds
when not in use, and provides approximately 30 hours
of use with the use of alkaline batteries.
Name and Model
Name: John Bunn DigiOx Finger Pulse Oximeter
Model: JB02008
BATTERY INSTALLATION
1. Refer to Figure 3; insert
two AAA batteries, oriented
as shown, into the battery
compartment.
2. Replace the cover.
Note: Ensure that the batteries
are correctly installed. Incorrect
installation will prevent the
device from operating.
OPERATION INSTRUCTIONS
Notes:
Clean the test finger with isopropyl alcohol before and
after each measurement. Ensure that the finger is dry
before being inserted into the oximeter.
Do not shake the finger; keep finger motionless during
oximeter use.
Avoid placing the device on a limb which is wrapped
with a cuff for blood pressure measurement or undergo-
ing venous infusion.
Ensure that no obstruction (contamination, scar tissue,
fingernail polish, acrylic fingernail, etc.) exists at the
site where the sensor is placed; such obstruction could
affect the signal received by the sensor and result in an
incorrect measurement.
Vigorous exercise and electrosurgical device interfer-
ence may affect the measuring accuracy.
1. Open the clip as shown in
Figure 4.
2.
Insert finger, nail up, into
the clip's rubber cushion
(ensure that the finger is
correctly positioned), and then gently clip the finger.
3. Press the Power/Display Button to turn the oxim-
eter on.
4. The screen will then display the
data as shown in Figure 5. After
the values become stable, the
user can read the information
on the display screen.
5. When the screen displays as shown in Figure 5,
pressing the Power/Display Button will change the
screen's orientation:
a.
Press the Power/Display Button
once more; the screen will
rotate 180° and display as
shown in Figure 6.
b. Press the Power/Display
Button once more; the screen
will rotate 90° and display as
shown in Figure 7.
c. Press the Power/Display Button
once more; the screen will
rotate 180° and display as
shown in Figure 8.
d. Press the Power/Display Button
once more; the screen will
rotate -90° and also show pulse
as a filled wave-form bar graph
along bottom of screen, as
shown in Figure 9.
e. Press the Power/Display Button
once more; the screen will now
show pulse as a hollow wave-
form bar graph along bottom of
screen, as shown in Figure 10.
f. Press the Power/Display Button once more; the
screen will return to the original screen as shown
in Figure 5 on previous page.
6. To change the oximeter’s display screen bright-
ness, press the Power/Display Button for more
than one second. There are 10 brightness levels,
and the level, from “BR 1” to “BR 10”, will show
on the screen during brightness adjustment. The
default is level four. Keep the Power/Display Button
depressed for as long as it takes to cycle to the
desired brightness level.
7. The oximeter will automatically turn off if no finger
is sensed for 8 seconds. If the finger is removed
and the Power/Display Button is pressed, the
oximeter will remain on and change screens, but
will display “FINGER OUT”.
ACCESSORIES (INCLUDED)
1 One lanyard
2 Two AAA alkaline batteries
3 One operation manual
REPAIR, MAINTENANCE, AND DISPOSAL
WARNING:
DO NOT open the oximeter housing.
It contains no serviceable parts. Service must be
performed by trained, authorized personnel only. Oth-
erwise, device failure and health hazard may occur.
Maintenance
In order to ensure the oximeter's longevity and
continued dependable operation, please observe the
following maintenance recommendations.
Change the batteries when the low-voltage indicator displays.
Clean the surface of the device before using: Wipe the device
with a clean, soft cloth saturated with a solution of 70% isopropyl
alcohol, then let it air dry or wipe it dry. Repeat as necessary.
Take out the batteries if the oximeter will not be used for a
long period of time.
Environment (Note:
Humidity may shorten
the life of the device,
or even damage it)
Operation: 41°F - 104°F
(5°C - 40°C), 15% - 80% RH
Storage: -4°F - 158°F
(-20°C - 70°C), < 93% RH
Cleaning and Disinfection
1. Do not immerse the device in liquid. Clean the de-
vice only by using a surface-clean technique: Wipe
with a clean, soft cloth saturated with a solution of
70% isopropyl alcohol.
2. Wipe with a clean, soft cloth saturated with clean
water and air dry, or dry with a clean, soft cloth.
Caution: Do not sterilize this device. Do
not subject this device to irradiation, steam,
ethylene oxide, high temperature, or immersion
in liquid.
Caution: Do not use the sensor if it is dam-
aged.
Disposal
Do not treat this device or its batteries as household
waste. Separate collection is required.
TROUBLESHOOTING
Problem Possible Reason Solution
SpO
2
and
Pulse Rate
display
unstable or
hard to see
Finger
incorrectly
positioned in
device
Reposition nger
correctly and try again
Finger is
shaking or
patient is
moving
Ensure patient remains
immobile
Patient’s SpO
2
value is too low
to measure
Remeasure; if
unsuccessful, yet device
is functioning correctly,
contact a medical
professional immediately
Excessive
ambient light
Reduce ambient light
Troubleshooting continued
Problem Possible Reason Solution
The device
does not
turn on
Batteries are
low
Replace batteries
Batteries
incorrectly
installed
Reinstall batteries
correctly
The device has
malfunctioned
Contact your GF Health
Products, Inc. distributor
The device
turns off
suddenly
The device
powers off
automatically
8 seconds
after nger is
removed
Normal
Batteries are
almost drained
Replace batteries
Screen
displays
“Error3”,
“Error4”,
“Error6” or
“Error7”
Device
malfunction
Contact your GF Health
Products, Inc. distributor
SYMBOL KEY
Symbol Description
Type BF
Warning: See operation manual
The pulse oxygen saturation
Pulse rate (beats per minute)
Low battery voltage
SN
Serial number
Not for continuous monitoring
TECHNICAL SPECIFICATIONS
Display screen Organic LED Display
Power supply
requirement
Two 1.5V AAA alkaline batteries
SpO
2
Parameter
Specications
Measuring range: 0%-100%
Accuracy: 70%-100%, ±3% < 70%,
no denition
Pulse Rate
Parameter
Specications
Measuring range: 30-235 bpm
Accuracy: 30-99bpm, ±2 bpm; 100-
235 bpm, ±2%
Dimensions (L x
W x H)
2.28" x 1.26" x 1.34" (58 mm x 32 mm
x 34 mm)
Environment Operation: 41°F - 104°F (5°C - 40°C),
15% - 80% RH
Storage: -4°F - 158°F (-20°C - 70°C),
< 93% RH
Net weight with
batteries
2 oz. (50g)
TECHNICAL SPECIFICATIONS continued
Classication The type of protection against electric
shock: Internally powered equipment
The degree of protection against
electric shock: Type BF applied part
The degree of protection against
harmful ingress of liquids: Ordinary
equipment without protection against
ingress of water
Electro-Magnetic Compatibility: Group
I, Class B
LIMITED WARRANTY
GF Health Products, Inc. warrants the John Bunn
DigiOx Finger Pulse Oximeter, Model JB02008, for a
period of one year against manufacturer's defects. If
a product is deemed to be under warranty, GF Health
Products, Inc. shall provide, at its option, (1) replace-
ment of any defective part or product or (2) a credit of
the original selling price made to GF Health Products,
Inc.'s initial customer. The warranty does not include
any labor charges incurred in replacement part(s) in-
stallation or any associated freight or shipping charges
to GF Health Products, Inc.
Graham-Field and John Bunn are registered
trademarks of GF Health Products, Inc.
Packaging, warranties, products, and specifications
are subject to change without notice.
GF Health Products, Inc. is not responsible for
typographical errors.
USA Corporate Headquarters:
GF Health Products, Inc.