Fluke 80K-6 Thermometer User Manual


 
LIMITED WARRANTY & LIMITATION OF LIABILITY
Each Fluke product is warranted to be free from defects in material and
workmanship under normal use and service. The warranty period is one
year and begins on the date of shipment. Parts, product repairs and
services are warranted for 90 days. This warranty extends only to the
original buyer or end-user customer of a Fluke authorized reseller, and
does not apply to fuses, disposable batteries or to any product which, in
Fluke’s opinion, has been misused, altered, neglected or damaged by
accident or abnormal conditions of operation or handling. Fluke warrants
that software will operate substantially in accordance with its functional
specifications for 90 days and that it has been properly recorded on non-
defective media. Fluke does not warrant that software will be error free or
operate without interruption.
Fluke authorized resellers shall extend this warranty on new and unused
products to end-user customers only but have no authority to extend a
greater or different warranty on behalf of Fluke. Warranty support is
available if product is purchased through a Fluke authorized sales outlet
or Buyer has paid the applicable international price. Fluke reserves the
right to invoice Buyer for importation costs of repair/replacement parts
when product purchased in one country is submitted for repair in another
country.
Fluke’s warranty obligation is limited, at Fluke’s option, to refund of the
purchase price, free of charge repair, or replacement of a defective
product which is returned to a Fluke authorized service center within the
warranty period.
To obtain warranty service, contact your nearest Fluke authorized service
center or send the product, with a description of the difficulty, postage and
insurance prepaid (FOB Destination), to the nearest Fluke authorized
service center. Fluke assumes no risk for damage in transit. Following
warranty repair, the product will be returned to Buyer, transportation
prepaid (FOB Destination). If Fluke determines that the failure was
caused by misuse, alteration, accident or abnormal condition of operation
or handling, Fluke will provide an estimate of repair costs and obtain
authorization before commencing the work. Following repair, the product
will be returned to the Buyer transportation prepaid and the Buyer will be
billed for the repair and return transportation charges (FOB Shipping
Point).
THIS WARRANTY IS BUYER’S SOLE AND EXCLUSIVE REMEDY AND
IS IN LIEU OF ALL OTHER WARRANTIES, EXPRESS OR IMPLIED,
INCLUDING BUT NOT LIMITED TO ANY IMPLIED WARRANTY OF
MERCHANTABILITY OR FITNESS FOR A PARTICULAR PURPOSE.
FLUKE SHALL NOT BE LIABLE FOR ANY SPECIAL, INDIRECT,
INCIDENTAL OR CONSEQUENTIAL DAMAGES OR LOSSES,
INCLUDING LOSS OF DATA, WHETHER ARISING FROM BREACH OF
WARRANTY OR BASED ON CONTRACT, TORT, RELIANCE OR ANY
OTHER THEORY.
Since some countries or states do not allow limitation of the term of an
implied warranty, or exclusion or limitation of incidental or consequential
damages, the limitations and exclusions of this warranty may not apply to
every buyer. If any provision of this Warranty is held invalid or
unenforceable by a court of competent jurisdiction, such holding will not
affect the validity or enforceability of any other provision.
Fluke Corporation Fluke Europe B.V.
P.O. Box 9090 P.O. Box 1186
Everett, WA 98206-9090 5602 BD Eindhoven
U.S.A. The Netherlands
5/94
®
80K-6
High Voltage Probe
Instruction Sheet
Description
The 80K-6 is a high voltage probe designed to extend the voltage
measuring capability of an ac/dc voltmeter to 6000 volts peak AC
or DC Overvoltage Category I. This means the probe can only be
used to make measurements on energy limited circuits within
equipment. Examples include high voltage within televisions or
photo copy machines. DO NOT use this probe to measure high
voltages on power distribution systems. A 1000:1 voltage divider
provides the probe with a high input impedance. The divider also
provides high accuracy when used with a voltmeter having a 10
megohm input impedance. A molded plastic body houses the
divider and protects the user from the voltage being measured.
Specifications
The 80K-6 will achieve rated accuracy when it is used with an
ac/dc voltmeter having 0.25% accuracy, or better, and an input
impedance of 10 megohms ±10%. Specifications for the probe
are as follows:
Voltage Range: 0 to 6 kV, dc or peak ac
Input Impedance: 75 megohms nominal
Division Ratio: 1000:1
Accuracy:
DC to 500 Hz: ±1%
500 Hz to 1 kHz: ±2%
Above 1 kHz: Output reading falIs.
Typically, -30% at 10 kHz.
Temperature:
Storage: -20 to 60°C
Operating: 0 to 50°C
Temperature Coefficient: 100 ppm/C
Relative Humidity: 20 - 80%
Altitude: 2000 m
Shock and Vibration:
Shock: 1 m Drop
Vibration: 5-55 Hz swept sine 3 g max.
Dimensions: 24.89 cm (9.8 in) (probe body), 5.10
cm (2.0 in) max. width
Weight: 12.7 g (45 oz)
Safety:
CAN/CSA C22.2 No. 1010.2.031-94 & IEC 1010-2-031:1993,
Type B, 6 kVDC or peak AC, 4.24 kV rms AC, Overvoltage
Cat. I (Voltage derived from limited energy transformers).
Measurement Considerations
Before attempting to use the 80K-6, the following paragraphs
should be read and understood. Particular attention should be
given to Operator Safety.
Operator Safety
Indicates the operator must refer to an explanation in
this manual.
Indicates terminals at which lethal voltages may exist.