SYSTem:CONFigure
This command sets the overall operating mode for the Agilent 6800-Series AC Power Source/Analyzers.
The choices are normal mode, which causes the product to operate as a standard ac source, or IEC mode,
which modifies the basic behavior of the transient and measurement systems to facilitate IEC
measurements. SYSTem:CONFigure has a variety of global consequences that are summarized below:
NORMAL MODE IEC MODE
Base Sampling Rate
39.920792 kHz 38.400000 kHz
Output Frequency
DC - 1000 Hz 50 Hz & 60 Hz Only
Freq/Window/Fs Mode
Independent Coupled
Transient System
FIXEd/STEP/PULSe/LIST
Modes
FIXEd Mode Only
Slew Operation
AC & DC Voltage; Frequency AC Voltage Only
MEASure:ARRay:CURRent
:HARM?
DC, Fundamental, and
Harmonics to 50th
Fundamental and Harmonics
to 40th plus RMS Current,
RMS Voltage & Real Power
CALCulate:SMOothing
Not/Available 1.5 Second Smoothing on/off
MEASure:ARRay:VOLTage
:FLUCtuations
Not/Available FLICker | PST | ALL?
CALCulate:INTegral:TIME
Not/Available 1 | 5 | 10 | 15 MINutes
Transmission of a SYSTem:CONFigure command implies ABORt and terminates any transient or
measurement actions previously initiated.
Command Syntax
SYSTem:CONFigure <CRD>
Parameters
NORMal | IEC
*RST Value
NORMal
Examples
SYST:CONF NORM
Query Syntax
SYSTem:CONFigure?
Returned Parameters
<CRD>
Related Commands
ABORt MEAS:ARR:CURR MEAS:ARR:VOLT CALC:SMO
CALC:INT SENS:WIND