Short Form Report
The Short Form report includes summary information about the current test. This information fits on one page.
The following is a sample short form report of a Fluctuating Current Harmonics test:
IEC 1000-3-2/EN 61000-3-2 Fluctuating Current Harmonics Test
Date Performed: 06/15/97
Test Executed By: A.C. Testright
Company Name: True Test Laboratories
Device Under Test ID: Model 12345 Modular Power Supply
Test ID: 00123
Approved by: QA
Signature:________________________________________________________ Date:__________________
Final Test Result: FAIL
Settings and Test Conditions Compliant to the Standard: Yes
Test Equipment Used:
Agilent 6842A Harmonic/Flicker Test System with serial number 3440A-00101
Agilent 14761A HFTS software Version: A.00.05
Date Last Calibrated: June 6, 1997
Test Equipment Settings:
------------------------
Line Voltage: 230.00 V Current Measurement Range: High
Line Frequency: 50 Hz Measurement Delay: 10.0 seconds
Device Class: D Fluctuating Harmonics Test Duration: 2.00 minutes
Class Determination Pre-test Duration: 10.00 seconds
Overrides:
----------
Test Limit Source (Power Measurements/Statistics): Maximum
Power Overrides: None
Test Limit Overrides: None
Pre-test Results for Class Determination:
-----------------------------------------
Percent in Envelope: 100.0% Fundamental Current: 1.343 A
Class D Equipment?: Yes Real Power: 148.8 W
Voltage THD Out-of-Specification?: No Power Factor: 0.387
Maximum Power: 148.8 W Mean Power: 146.7 W
Total Number of Failures: Total Number of Errors:
------------------------- -------------------------
4505 Failures None
Remarks
-------