Setting Up the Test - 3
25
ñ Select the number of integration periods to be acquired during the test. The Resulting Time
field displays the test duration based on the number of integration periods as well as the Pst
integration time that you have specified in the Advanced window. The Pst integration time is set
to a default of 10 minutes with the Advanced options disabled.
Advanced Test Options
ñ Select the rms and peak current limit of the Agilent 6800-Series AC Power Source/Analyzer that
you will be applying to the equipment under test. Generally, it is best to leave these parameters
set to their maximum default values. Although the function of the current limit is to protect the
equipment under test, if the unit does go into current limit during a normal test, the test will be
invalid.
ñ Select the returned data type (All, Pst/Rms Summary, or Rms/Flicker Time Series)
ñ Select the Pst (or short-term) flicker integration time.
ñ Select a test limit override for the Pst/Plt Thresholds (short-term/long-term flicker) and Rms
Thresholds. This is specified as a percent of the specification in the applicable regulation.
ñ Select a measurement delay from the time power is applied to the equipment under test until the
time that the Agilent 6800-Series AC Power Source/Analyzer starts measuring data.